1. Patient Preparation
1.1 Pre-Scan Requirements
► Fasting: 4-6 hours (water permitted) to stabilize blood glucose levels
► Hydration: 500mL water intake 1 hour pre-scan to enhance radiotracer clearance
► Contraindications Check: Pregnancy, uncontrolled diabetes (blood glucose >200mg/dL), severe claustrophobia
1.2 Radiopharmaceutical Administration
► FDG Dose: 0.1-0.15 mCi/kg (3.7-5.5 MBq/kg) via IV catheter
► Uptake Period: 60±5 minutes in quiet, warm environment (reduce brown fat uptake)
► Bladder Voiding: Mandatory pre-scan to minimize pelvic artifact
2. Equipment Setup
2.1 System Preparation
► Daily QC: Ge-68 phantom calibration & CT dose verification
► Patient Positioning:
-
Arms raised for whole-body scans (reduces beam hardening)
-
Head first supine for brain studies (use head holder)
2.2 Protocol Selection
Examination Type | CT Parameters | PET Acquisition |
---|---|---|
Oncology WB | 120kV, 50mAs | 2min/bed, 6-7 beds |
Brain Neurological | 80kV, 30mAs | 10min static |
Cardiac | ECG-gated 100kV | List-mode 20min |
3. Scanning Process
3.1 CT Acquisition
► Scout View: Topogram (AP/LAT) for scan range planning
► Low-Dose CT: For attenuation correction (30-50mAs)
► Diagnostic CT: Optional contrast-enhanced (portal venous phase)
3.2 PET Acquisition
► 3D Mode: Standard for most clinical applications
► Time-of-Flight (TOF): Enabled for BMI>30 patients
► Respiratory Gating: For lung/liver lesions (optional bellows)
4. Post-Processing & Analysis
4.1 Image Reconstruction
► Iterative Algorithm: OSEM (2 iterations, 20 subsets) + PSF correction
► Matrix Size: 200×200 for body, 256×256 for brain
4.2 Quantification
► SUV Calculation: Lean body mass normalization (SUVlbm)
► Lesion Analysis: Max/mean SUV, metabolic tumor volume (MTV)
5. Radiation Safety Protocol
► Staff Dosimetry: Wear TLD badges, maintain >2m distance during FDG injection
► Patient Clearance: Confirm SUV <2.0 in bladder before discharge
► Contamination Check: Wipe test for injection area
6. Brand-Specific Notes
-
Siemens Biograph: Enable UltraHD-PET for <2mm resolution
-
GE Discovery: Use Q.Clear for Bayesian penalized reconstruction
-
Philips Vereos: Digital photon counting requires daily detector calibration
Scan Time Optimization:
▸ Whole-body (skull-to-thigh): 15-20 minutes total
▸ Brain/Cardiac: 25-30 minutes